Sorry, you need to enable JavaScript to visit this website.
hamburger overlay

Frequently Asked Questions

Cleaning the meter

Be sure to clean the areas around slots and openings, while being careful not to get any moisture in them. You should also wipe down the entire meter surface, including the meter display.

  • 70% isopropyl alcohol
  • Mild dishwashing liquid mixed with water
  • 10% household bleach solution (1 part bleach plus 9 parts water) made the same day

Special Features

 

  

 

 

Accu-Chek Guide meter can store up to 720 test results with their time and date. You do not need to do anything to save the results. The meter automatically saves all test results.
If all the memory locations are occupied, the oldest test result is deleted to create space for the new one when you perform a new test.

Yes, you can set up to 4 general test reminders per day and switch on the post-meal reminder in the set-up mode.

 1.   After performing a blood glucose test, the test   result is displayed on the screen with Add   Comment highlighted. Press  to add a   comment.
 2.  The Add Comment menu appears. Press  to   highlight the desired comment (the example here   is Before meal). Press  to select the comment.
 3.

 If Patterns is On:

 If Patterns is on and you select Before meal   or After meal, press  to select a specific meal   (Breakfast, Lunch, Dinner, or Snack). Press  to   set the selected meal for the test result.

 4.
 The final result appears with the comment symbol.   Press  or  to set the comment and return   to Main Menu. To change the comment, press  to select the comment. Press  to return to Add   Comment.

Blood Glucose Monitoring

No, check the use by date  on the test strip container. Do not use test strips past the use by date. 

A ‘Drop Too Small’ (E4) error message will display on the meter, indicating not enough blood was drawn into the test strip for measurement or blood was applied after the test had started. You should discard the test strip and repeat the blood glucose test.

Only a tiny blood drop will do (0.6 microlitre).

  1. Insert your test strip into the meter and it turns on automatically.
  2. Apply your blood sample to the easy to see yellow dosing area. Only 1 small drop is needed.
  3. In less than 4 seconds, your test results will appear in the easy to read display.
Still can't find what you need?

We want your feedback!

GO

Accu-Chek Guide

Item Description

Wireless connection to your compatible smartphone keeps your important information in the palm of your hand, eliminating the need for a paper diary. Along with proven accuracy1, features like the spill resistant vial, wide test strip dosing area, strip port light and strip ejector button can help make every day blood glucose monitoring easier.2

References:
1. Accu-Chek Guide Test Strip Package Insert 2016.
2. Harvey, C., et al., J Diabetes Sci Technol 2016 10(5):1136-1142.

Product Manuals

DISPLAY

LCD

RESULTS

Less than 4 seconds

EASY

No coding

SAMPLE

Tiny: 0.6µL

WEIGHT

40g with battery

Benefits and Features
  • Proven accuracy that exceeds current ISO 15197:2013
  • Have the option of transmitting results to the Accu-Chek Connect diabetes management app on your compatible smartphone using wireless technology
  • No more "strip confetti"—the spill resistant vial helps you remove one strip at a time
  • Wide test strip dosing area —apply a small drop of blood anywhere along the wide yellow edge
  • Receive glucose readings in less than 4 seconds
  • Monitor in all light conditions, thanks to a bright, backlit display and a strip port light
  • Eject used test strips from the meter with the push of a button
  • Follow step by step, written instructions on the display
  • Enjoy virtually pain free2 lancing with the included Accu-Chek FastClix lancing device

References:
1. Accu-Chek Guide Test Strip Package Insert 2016.
2. Kocher, S., et al., J Diabetes Sci Technol 2009 Sep 1; 3(5):1136-1143

Additional Specifications
Measuring principle:

FAD glucose dehydrogenase (GDH)

Measuring time:

Less than 4 seconds

Measuring conditions:

Temperature: 4°C to 45°C 

Memory capacity:

720 blood glucose results and 32 control tests results with time and date

Test averages:

7, 14, 30 and 90 days 

Data transfer:

Bluetooth (with mobile devices)
Micro USB cable (with computer)

Dimensions:

80 x 47 x 20 mm 

Weight:

40g with battery 

Display:

LCD

Beep tone:

Yes (factory setting ON) 

Battery:

2 battery, 3V type 2032 

Auto power off:

90 seconds

Under dosing detection:

Yes

Altitude independence:

0 to 3,094 m (0 to 10,150 ft.)

Sample size:

0.6 µL 

Measuring range:

0.6 mmol/L – 33.3 mmol/L 

Sample types:

Capillary, venous, arterial, and neonate

Sample dosing:

Capillary fill function

Hematocrit range:

10 to 65% 

Test strip stability:

Test strips remain stable up to expiry date printed on test strip vial, even after opening (test strip container must be closed tightly after each test strip is removed)

How to Use

Get results in 3 steps:

  1. Insert your test strip into the meter and it turns on automatically.
  2. Apply your blood sample to the easy to see yellow dosing area. Only 1 small drop is needed.
  3. In less than 4 seconds, your test results will appear in the easy to read display.