Frequently Asked Questions

200 units

The device should be disposed in accordance with the standards set by your institution.

With children younger than 1 year of age, the capillary blood must be obtained from the heel. Obtaining blood must occur from the lateral areas of the sole.

The puncture must occur vertically to the sole. While obtaining blood, the child should be held with its foot vertically facing the floor. A site which has already been punctured must not be punctured a second time.

To improve blood flow at the heel before obtaining blood, you can carefully warm the heel, for example by wrapping it in a warm, moist cloth of a maximum 42 °C.

Yes.

No. The lancing device is designed and engineered for one-time use.

Choose the low setting for soft, thin and delicate skin or use on average skin to obtain a smaller drop of blood.

Choose the medium setting for normal skin or if a medium blood volume is required.

Choose the high setting for calloused skin, to obtain a blood sample without pressing and squeezing, if a high blood sample is required or for multi-test situations.

Medium depth setting

Still can't find what you need?

We want your feedback!

GO

Accu-Chek Safe-T Pro Plus

Item Description

Multiple safety features combined with customised settings make the Accu-Chek Safe-T Pro Plus lancing device especially advantageous. This single-use lancing device supports patient safety and ease-of-use. It also complies with the safety requirements demanded by health authorities, legislators, and health institutions.

Product Manuals

DEPTH

1.3, 1.8 and 2.3 millimeters

DISPOSAL

Entire device is for single-use only. Disposal of used single-use lancing devices always in accordance with the local / facility’s disposal procedures

INTENDED USE

Single-use, obtaining capillary blood from the fingertip or, if the patient is a child under 1-year, from the heel

Benefits and Features
  • 3 adjustable depth settings allow the sampling of blood according to the skin condition of the patient
  • Each depth setting clicks into place and cannot slip out of position once selected
  • Special beveled needle cut and needle diameter for almost pain-free blood sampling
  • Suitable for adults, children under 1-year and neonates
  • Save space and costs by using one type of lancet
  • Available in a box with 200 single-use lancing devices
  • Designed for single-use, to avoid cross contamination
  • The Safe-T technology ensures that the needle is always safely stored in the housing, preventing accidental needle sticks
  • Automatic needle retraction after single-use eliminates risk of injury and cross contamination
  • After the single-use the complete lancing device is disposed of
Additional Specifications
Intended use:

Single-use only. Obtaining capillary blood from the fingertip or, if the patient is a child under 1-year, obtain from the heel

Adjustable depth setting:

3 adjustable depth setting

Penetration depth:

Low (short marking) – 1.3mm
Medium (middle marking) – 1.8mm
High (long marking) – 2.3mm

Needle diameter:

0.63 mm (23 G)

Special needle cut:

3 facet cut

Siliconized needle:

Yes

Automatic retraction of needle:

Yes

Sterility:

Till expiry date printed on box

How to Use

 

Steps Picture Guide
 1. Twist off the sterility cap in either             direction. 
 2. Select the depth you want by turning       the depth adjuster (Preset at medium       depth).
 3. Press the lancing device firmly                 against the puncture site, push the           relaesse and collect blood.

 